Servo-i Neonatal NAVA Management Protocol
This document outlines the procedure for placing and positioning the Edi catheter for Neonatal NAVA (Neurally Adjusted Ventilatory Assist) management. It details the specific steps for connecting the Edi module, selecting the correct catheter size based on patient weight and height, and verifying catheter position using waveform analysis and color indicators. The guide also specifies the required ventilator settings for both invasive and non-invasive NAVA modes on the Servo-i ventilator.
Covers: Catheter Placement and Positioning · Ventilator Settings · Alarm Settings · Troubleshooting Guide · Assessing Edi · Catheter Information
Latest Servo-i manuals
Servo-i Neonatal NAVA Management Protocol
User Manual 7 pages · 230 KB · ENA 2021 protocol for catheter placement, ventilator settings, and troubleshooting for Neonatal NAVA management on the Servo-i ventilator.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in Servo-i manuals
How do I select the correct Edi catheter size?
Select the catheter size based on the patient's weight and height using the provided chart (e.g., 6 Fr/49cm for 500g—1.5kg).
What indicates the Edi catheter is properly positioned?
Proper positioning is indicated by a diminishing ECG waveform progressing from the 1st to the 4th waveform and the presence of a pink color in the 2nd and 3rd waveforms.