This user manual provides detailed instructions for installing, operating, and calibrating the Digital AVC vacuum gauges (DAVC-4, DAVC-5, DAVC-6). It covers power requirements, serial and analog output configurations, alarm settings, and calibration procedures using reference tubes. The manual also includes safety information, compliance data, and accessory details for compatible gauge tubes and extension cables.
The Digital AVC is a digital readout version of THI’s AVC vacuum gauge, using heated noble-metal thermocouple tubes for pressure measurement. It supports three tube families (DV-4, DV-5, DV-6) with different pressure ranges and provides both digital and analog outputs for integration with control systems.
| Input Power | 11 – 32 VDC |
|---|---|
| Weight | 0.27 lbs (123 Grams) |
| Operating Temperature Range | -20°C to 70°C |
| Accuracy (DV-4) | ± (20% of Reading + 0.01 Torr) |
| Accuracy (DV-5) | ± (20% of reading + 0.2 mTorr) |
| Accuracy (DV-6) | ± (15% of Reading + 0.001 mTorr) |
| Digital Output | RS232 (9600 / 19.2k baud) |
| Tube Leak Test | <1 x10-8 atm cc/sec |
DAVC-4: 0.2 - 20 Torr; DAVC-5: 0.1 – 100 mTorr; DAVC-6: 1.0 - 1000 mTorr.
The DAVC requires 11–32 VDC, with a minimum power draw of 0.7 Watts.
The DV-6 tube has a thermal coefficient of 0.2 mTorr/°C + 0.25% of reading/°C.