This is a Microvision2 hardware manual
This manual provides detailed instructions for installing, maintaining, and operating the Microvision2 RGA control electronics package. It covers mechanical and electrical specifications, rear panel connections, analyzer installation, baking procedures, and ion source maintenance. Safety warnings and compliance with European regulations are included. The device is designed for use with MKS software and TCP/IP communication, and requires proper grounding and ventilation during operation.
The manual also outlines procedures for returning equipment for service, including required documentation and packaging instructions. It emphasizes that maintenance should only be performed by qualified personnel and that no user-serviceable parts exist within the unit.
| Dimensions | 127mm wide x 127mm x 179mm deep, plus 46mm over analyser connector to face of CF flange |
|---|---|
| Weight | 1.7 kg |
| Power Supply Voltage | 100 to 120, 200 to 240 VAC rms |
| Operating Temperature | 0 to 40°C |
| Max Operating Pressure | 7.6x10 -5 (1x10 -4 mbar) |
The maximum operating pressure is 7.6x10 -5 (1x10 -4 mbar).
The power supply accepts 100 to 120, 200 to 240 VAC rms at 47-63Hz @ 2.1A rms.
Yes, it can be operated from an IBM compatible PC or MAC running MKS Process Eye Professional or EasyView software.
No, there are no user-serviceable parts in the electronic equipment.