This is a Laboratory Nitrogen Generator G1-G9 user guide, which mainly introduces the use, installation and maintenance methods of the nitrogen generator
The manual provides comprehensive instructions for the safe handling, unpacking, and installation of Domnick Hunter Laboratory Nitrogen Generators (models G1-G9). It details technical specifications including flow rates, purity levels, and physical dimensions for various configurations with or without integral compressors.
Key operational topics include start-up procedures, generator functions such as Economy Mode and Rapid Cycle, and specific bypass assembly instructions for G4 models. The guide also outlines fault-finding procedures for common issues related to inlet/outlet connections and electrical faults.
| Flowrate | Up to 25.0 L/min (G4) |
|---|---|
| Input Voltage | 115/230V ac 50/60Hz ± 10% |
| Operating Temperature | 5 – 45 oC (41 – 113) |
The generator should be run for a minimum of six hours prior to use with the application.
In the event of any kind of interruption, the unit must be run for 1-hour offline to ensure purity.
Economy mode stops cycling and turns off valves if backpressure exceeds a limit, displaying 'Economy' until demand returns.