Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines

This document introduces the testing method of high level disinfectant Revital-Ox RESERT™, including the required reagents, temperature and time.


Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines - cover page
Brand
Steris
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
162 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
6EEE5FB1006A0C3E50698036669B00EB

About this manual

This document provides guidelines for Endoscopy Staff (RN, LPN, or Endo Tech) on how to verify that the concentration of hydrogen peroxide in Revital-Ox RESERT™ High Level Disinfectant is above the minimum recommended concentration (MRC) of 1.5%. The procedure involves using specific test strips and a timer to ensure the solution remains effective for disinfecting semi-critical equipment.

What's inside

  • Supplies needed
  • Confirming minimum recommended concentration
  • Quality Control of Test Strips
  • PASS solution preparation
  • FAIL solution preparation
  • Interpretation of results

Frequently asked questions

What is the minimum recommended concentration for Revital-Ox RESERT™?

The minimum recommended concentration (MRC) is 1.5% hydrogen peroxide.

How often should the disinfectant solution be tested?

The solution must be tested prior to each use.

What indicates a PASS result on the test strip?

A PASS result occurs if the pad is completely blue or purple.

How long should a test strip remain in the solution?

The strip should be dipped for 2 seconds and then removed to start a 60-second timer.