Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines
This document provides guidelines for Endoscopy Staff (RN, LPN, or Endo Tech) on how to verify that the concentration of hydrogen peroxide in Revital-Ox RESERT™ High Level Disinfectant is above the minimum recommended concentration (MRC) of 1.5%. The procedure involves using specific test strips and a timer to ensure the solution remains effective for disinfecting semi-critical equipment.
Covers: Supplies needed · Confirming minimum recommended concentration · Quality Control of Test Strips · PASS solution preparation · FAIL solution preparation · Interpretation of results
Latest Steris manuals
Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines
User Manual 2 pages · 162 KB · ENGuidelines for endoscopy staff to verify the hydrogen peroxide concentration of Revital-Ox RESERT high-level disinfectant using test strips.
Which document do I need?
User Manual
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in Steris manuals
What is the minimum recommended concentration for Revital-Ox RESERT™?
The minimum recommended concentration (MRC) is 1.5% hydrogen peroxide.
From: Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines
How often should the disinfectant solution be tested?
The solution must be tested prior to each use.
From: Steris Revital-Ox RESERT High Level Disinfectant Testing Guidelines