About this manual

The DT722-X18 (Analog) is a rugged, industrial relative humidity and temperature transmitter designed for process applications requiring accurate measurement and control. It features a stainless steel housing and can withstand operating temperatures up to 150°C (300°F). The device offers long-term stability of ±1% RH over 12 months and utilizes an M12 fast connector for electrical connections.

What's inside

  • Introduction
  • Features
  • Installation
  • Electrical Connections
  • Calibration and Adjustment
  • Safety
  • Toxic Materials
  • Repair and Maintenance
  • Technical Specifications
  • Recommended Practices

Specifications

Operating Temperature Range0 to 150°C (32 to 300°F)
Operating Humidity Range0 to 100% RH
Long Term Stability±1% RH over 12 months
Connector TypeM12 fast connector
Housing MaterialStainless steel

Frequently asked questions

How often should the instrument be calibrated?

An annual calibration is recommended for an accuracy requirement of ±2% RH under ambient conditions.

What is the recommended calibration interval for ±5% RH accuracy?

A calibration is recommended every five years for an accuracy of ±5% RH.

Can the P3, P4, P6, P7, and P8 potentiometers be adjusted by the user?

No, these potentiometers are factory-set and must not be adjusted as it will affect instrument performance.