This document describes the features, options, and ordering codes for the Michell QMA601 process analyzer base unit and sampling system.
This document provides a feature description for the Michell QMA601 Process Analyzer, detailing various configuration options for the base unit and sampling system. It covers certifications such as ATEX/IECEx and cQPSus, measurement ranges from low to high, and different calibration methods including standard and asymmetric calibration.
The guide also outlines specifications for the sampling system, including imperial and metric fittings, indoor and outdoor enclosures, fast loop bypass configurations, heating options, and material certifications. It includes ordering examples and information on accessories, spare parts, and service packages.
| Certification (Base Unit) | ATEX/IECEx certified system |
|---|---|
| Low Range Measurement | ~0.5 ppmv |
| Medium Range Measurement | ~5 ppmv |
| High Range Measurement | ~50 ppmv |
| Standard Calibration Range | 0.1 ppmv to 700ppmv |
| Input Pressure (Standard) | 2 barg input |
| Back Pressure (Standard) | 1 barg back pressure |
| Sampling System Fitting (Imperial) | 1/4" tube and fittings |
| Sampling System Fitting (Metric) | 6mm tube and fittings |
| Indoor Sampling Enclosure | 316 stainless steel panel |
| Outdoor Sampling Enclosure (304) | 304 stainless steel enclosure |
| Outdoor Sampling Enclosure (316) | 316 stainless steel enclosure |
| Heater Voltage (240VAC Fixed) | 20°C |
| Heater Voltage (110VAC Fixed) | 20°C |
| Heater Voltage (Adjustable) | 0-60°C |
| Trace Heating Length | 3m |
| Trace Heating Material | 1/8" O.D. AISI 316 |
| Material Certificate Standard | BS EN 10204 – type 3.1 |
| Max Trace Heated Sample Line | 15m |
The standard calibration covers a range from 0.1 ppmv to 700ppmv.
This heater option is not for use with the F7 enclosure cooling feature.
The maximum length is 15 meters.
The maximum ambient temperature is +35°C.