Rotronic RMS-TCD-S digital temperature probe is a leading-edge product that meets all established standards. It will work perfectly for years if you follow the assembly and operating guidelines below.
This document is a short instruction manual for the Rotronic RMS-TCD-S-001 digital temperature probe. It provides general guidance on connecting the probe to a transmitter or cable, ensuring proper alignment and tightening for IP protection. The manual also outlines mounting requirements, suggesting complete immersion in the environment to be measured and noting that temperature mapping may be needed for regulated applications.
| Order code | RMS-TCD-S-001 |
|---|---|
| Measurement range | -200...200°C |
| Accuracy | ±0,25 °C bei -50…85 °C |
| Long term stability | ±0.5 °C/year |
| T63 | 90 s |
| Electronics temperature range | -40…85 °C |
| Relative humidity range | 0…100 %rh |
| Housing material | Polycarbonate |
| Housing diameter | Ø15 mm |
| Housing length | 110 mm |
| IP Rating (Housing) | IP65 |
| Connector type | Standard Rotronic 7-pin connector |
| Temperature sensor type | 4 wire PT100 1/3 DIN |
| Sensor material | Stainless steel - austenitic - 1.4571 (316Ti) |
| Sensor diameter | Ø3 mm |
| Sensor length | 50 mm |
| IP Rating (Sensor) | IP68 |
| Cable material | Perfluoroalkoxy alkanes (PFA) |
| Cable diameter | Ø 2.2 mm |
| Cable length | 2 m |
We recommend a calibration after 1 year of use, depending on the application.
You can perform calibration on-site with a temperature generator such as the HG2-S.
It is compatible with RMS-LOG-L, RMS-LOG-868, RMS-LOG-915, and RMS-LOG-L-D.