Rotronic RMS-TCD-S-001 Digital Temperature Probe Short Instruction Manual

Rotronic RMS-TCD-S digital temperature probe is a leading-edge product that meets all established standards. It will work perfectly for years if you follow the assembly and operating guidelines below.


Rotronic RMS-TCD-S-001 Digital Temperature Probe Short Instruction Manual - cover page
Brand
Rotronic
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
402 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
56AEFF5F4D07A8625DDD2747321A435F

About this manual

This document is a short instruction manual for the Rotronic RMS-TCD-S-001 digital temperature probe. It provides general guidance on connecting the probe to a transmitter or cable, ensuring proper alignment and tightening for IP protection. The manual also outlines mounting requirements, suggesting complete immersion in the environment to be measured and noting that temperature mapping may be needed for regulated applications.

What's inside

  • General Description
  • Connection
  • Mounting
  • Adjustment / Calibration
  • Technical Data

Specifications

Order codeRMS-TCD-S-001
Measurement range-200...200°C
Accuracy±0,25 °C bei -50…85 °C
Long term stability±0.5 °C/year
T6390 s
Electronics temperature range-40…85 °C
Relative humidity range0…100 %rh
Housing materialPolycarbonate
Housing diameterØ15 mm
Housing length110 mm
IP Rating (Housing)IP65
Connector typeStandard Rotronic 7-pin connector
Temperature sensor type4 wire PT100 1/3 DIN
Sensor materialStainless steel - austenitic - 1.4571 (316Ti)
Sensor diameterØ3 mm
Sensor length50 mm
IP Rating (Sensor)IP68
Cable materialPerfluoroalkoxy alkanes (PFA)
Cable diameterØ 2.2 mm
Cable length2 m

Frequently asked questions

How often should the probe be calibrated?

We recommend a calibration after 1 year of use, depending on the application.

What generator can be used for on-site calibration?

You can perform calibration on-site with a temperature generator such as the HG2-S.

Which transmitters is the RMS-TCD-S compatible with?

It is compatible with RMS-LOG-L, RMS-LOG-868, RMS-LOG-915, and RMS-LOG-L-D.