ROTRONIC MP101A Humidity Temperature Probe Instruction Manual

This manual introduces the use of mp 101a humidity and temperature probe, including use method, installation method, maintenance method, calibration method and other content.


ROTRONIC MP101A Humidity Temperature Probe Instruction Manual - cover page
Brand
Rotronic
Document type
User Manual
Language
English
Pages
11
File format
PDF
File size
45 KB
Published
16 October, 2023
Updated
16 October, 2023
MD5 checksum
D6A3E82E9D5A4270DC8F7EBABC186ED7

About this manual

The ROTRONIC MP101A is a combined humidity and temperature probe designed primarily for outdoor applications. It features low power consumption and linear output signals suitable for data acquisition systems. The device utilizes a capacitive humidity sensor with advanced filter technology to ensure exceptional resistance to contaminants.

The manual covers operation details, including power supply requirements (4.8-30 VDC) and output signal specifications for both relative humidity and temperature. It also provides detailed installation instructions regarding probe location, grounding recommendations, and wiring configurations for various connector types.

What's inside

  • Description
  • Operation
  • Installation
  • Maintenance
  • Calibration Basics
  • Specifications

Specifications

Supply Voltage4.8 and 30 VDC
Current Consumption10 mA
Energizing Time0.25 seconds
Operating Temperature Range-40 °C and +60 °C
Humidity Operating Range0 and 100 %RH
Output Impedance LimitLess than 1000 Ω

Frequently asked questions

How long does the probe need to be energized for measurement?

The probe must be energized for only 0.25 seconds during each measurement.

What are the operating temperature limits?

The MP101A can operate within -40 °C and +60 °C.

How often should the unit be calibrated?

For maximum accuracy, calibration of the unit may be verified every 6 to 12 months.