This manual introduces the operation of N-400 Fetal Pulse Oximetry Monitor, including warnings, cautions, notes, device description, routine maintenance, cleaning, spare parts, troubleshooting guide, packing for shipment, specifications, etc.
This document is the Technical Manual for the Nellcor Puritan Bennett N-400 Fetal Oxygen Saturation Monitor. It provides detailed information for qualified service personnel regarding the device's operation, routine maintenance, and troubleshooting. The manual describes the system components, including the monitor, patient module, and sensor, and explains features such as automatic calibration and audible alarms.
| Measurement Type | Functional oxygen saturation of arterial hemoglobin (FSpO2) and pulse rate |
|---|---|
| Update Rate | at each qualified pulse |
| Maintenance Interval | once every two years |
| Ground Wire Resistance Limit | ≤ 0.2 Ohms |
| Enclosure Leakage Current Limit | ≤ 100µAmps |
| Patient Leakage Current Limit (Normal) | ≤ 20µAmps |
Only qualified service personnel should service this product.
The default state of the audible alarm is OFF.
A fetal response to stimulation, such as vaginal exam or sensor insertion, can cause a reflex bradycardia.
The sensor is not making adequate contact at the sensor site on the fetus.