Nellcor Puritan Bennett N-400 Fetal Pulse Oximetry Monitor Technical Manual

This manual introduces the operation of N-400 Fetal Pulse Oximetry Monitor, including warnings, cautions, notes, device description, routine maintenance, cleaning, spare parts, troubleshooting guide, packing for shipment, specifications, etc.


Nellcor Puritan Bennett N-400 Fetal Pulse Oximetry Monitor Technical Manual - cover page
Brand
NELLCOR PURITAN BENNETT
Category
Professional medical equipment
Document type
Service Manual
Language
English
Pages
77
File format
PDF
File size
5.1 MB
Published
16 October, 2023
Updated
16 October, 2023
Model
N-400
MD5 checksum
4C97C4F39B82FD76FA0C2E883AFF53D3

About this manual

This document is the Technical Manual for the Nellcor Puritan Bennett N-400 Fetal Oxygen Saturation Monitor. It provides detailed information for qualified service personnel regarding the device's operation, routine maintenance, and troubleshooting. The manual describes the system components, including the monitor, patient module, and sensor, and explains features such as automatic calibration and audible alarms.

What's inside

  • Introduction
  • Warnings, Cautions, And Notes
  • Device Description
  • Routine Maintenance
  • Troubleshooting Guide
  • Packing For Shipment
  • Specifications
  • Theory Of Operation
  • Block Diagram Analysis
  • Schematic, Part Locator, and Block Diagrams

Specifications

Measurement TypeFunctional oxygen saturation of arterial hemoglobin (FSpO2) and pulse rate
Update Rateat each qualified pulse
Maintenance Intervalonce every two years
Ground Wire Resistance Limit≤ 0.2 Ohms
Enclosure Leakage Current Limit≤ 100µAmps
Patient Leakage Current Limit (Normal)≤ 20µAmps

Frequently asked questions

Who should service the N-400 monitor?

Only qualified service personnel should service this product.

What is the default state of the audible alarm?

The default state of the audible alarm is OFF.

What causes a reflex bradycardia during sensor placement?

A fetal response to stimulation, such as vaginal exam or sensor insertion, can cause a reflex bradycardia.

What should be done if the SENSOR LIFTED indicator remains ON?

The sensor is not making adequate contact at the sensor site on the fetus.