This document introduces the Puritan Bennett 840 ventilator information
This document serves as a quick reference pocket guide for the Puritan Bennett 840 Ventilator System. It provides an overview of the system components, including the Breath Delivery Unit (BDU) and Graphic User Interface (GUI), and explains how to connect the device to AC power, air, and oxygen supplies. The guide details the setup process for new patients, including specific settings for apnea and Bi-Level modes, and explains how to use the ventilator's alarm system to monitor patient status.
| Inspiratory flow | up to 200 L/min |
|---|---|
| Supply pressure | 35 to 100 psi |
| Safety ventilation pressure limit | 20 cm H2O |
| Idle flow | 5 L/min |
The 802 Backup Power Source operates the ventilator when AC power drops below a minimum level and recharges during AC power operation.
The ventilator enters idle mode, opens the exhalation valve, begins idle flow, and disables breath triggering.