This document introduces the software updates for Puritan Bennett™ 560 and Puritan Bennett™ 520 ventilators, including the addition of a respiratory trigger sensitivity level for pediatric patients, the ability to set respiratory rate and inspiratory time in assist/control ventilation mode, the provision of a circuit check function, the option to select auxiliary alarm sounds, and the ability to restore default ventilator settings.
This document serves as an appendix to the User Manual and Clinical Personnel Manual for the Puritan Bennett 560 and Puritan Bennett 520 ventilators. It details the enhancements provided by specific software updates (LX010101/02/03 for the 560 and LS010101/02/03 for the 520), which include new respiratory trigger sensitivity levels, expanded settings for respiratory rate and inspiratory time, and a new circuit leak check function.
The manual provides specific instructions and parameter ranges for these new features. Key additions include a new '0P' sensitivity level for pediatric patients, the ability to set respiratory frequency from 1-60 bpm in certain modes, and direct setting of inspiratory time in seconds. It also outlines the procedure for performing a 'Circuit Check' to detect leaks in the patient circuit, including safety precautions and interpretation of results. This manual is written in Chinese.
| Respiratory Rate Range (PA/C Mode) | 1-60 bpm |
|---|---|
| Respiratory Rate Range (VA/C Mode, PB560) | 1-60 bpm |
| Inspiratory Time Range (PA/C Mode) | 0.3-6.0 seconds |
| Inspiratory Time Resolution | 0.1 seconds |
| Circuit Check Pressure | 30 mbar |
| Leak Threshold for Failure | > 1 L/min |
A new level '0P' is added specifically for pediatric patients.
During startup, hold down the 'Menu' key to activate the circuit check function.