Fluke 787 ProcessMeter Calibration Manual

This manual introduces the use of Model 787 ProcessMeter Calibration, including product specifications, verification testing, maintenance, and calibration


Fluke 787 ProcessMeter Calibration Manual - cover page
Brand
Fluke
Category
Other Electronics
Document type
Service Manual
Language
English
Pages
39
File format
PDF
File size
735 KB
Published
15 October, 2023
Updated
15 October, 2023
MD5 checksum
A6165A4AA430368FD43C0FE377174C75

About this manual

The Fluke 787 ProcessMeter Calibration Manual is a technical guide designed for qualified personnel to perform verification tests and calibration adjustments on the 787 ProcessMeter. It details the necessary safety precautions, required equipment, and step-by-step procedures for verifying various measurement functions such as DC/AC volts, current, resistance, and frequency.

What's inside

  • Safety Information
  • Required Equipment
  • Specifications
  • Verification Tests
  • Calibration Procedures
  • Basic Maintenance
  • Replaceable Parts

Specifications

Overvoltage protection1000 V
Input impedance10 M Ω (nominal), < 100 pF
Operating Temperature+18 °C to +28 °C

Frequently asked questions

Who is qualified to perform the procedures in this manual?

The information provided in this manual is for the use of qualified personnel only.

What is the minimum required equipment for calibration?

A Fluke Model 5500A or 5520A Digital Multimeter is required as a calibration source.

Can the 787 Calibration Module be used during regular operation?

No, the 787 Calibration Module is not rated for Overvoltage Category III and should not be used during regular operation.