This document provides calibration information for the Fluke T+ and T+ PRO Electrical Testers. It also provides information on safety, symbols, contacting Fluke, specifications, maintenance, performance tests, parts and accessories list, and product warranty statement.
This manual serves as a reference guide for the Fluke T+ and T+ PRO Electrical Testers. It outlines critical safety information regarding electric shock hazards, proper usage in hazardous locations, and battery handling. The document details the technical specifications of the device, including operating temperature ranges, accuracy ratings for AC/DC voltage, resistance, and GFCI testing, as well as physical dimensions and weight.
| Operating Temperature | -10 °C to 55 °C (14 °F to 131 °F) |
|---|---|
| Storage Temperature | -30 °C to 60 °C (−22 °F to 140 °F) |
| Battery Type | AAA (2) |
| Maximum Working Voltage Earth Ground | 1000 V ac or dc |
| AC Voltage Accuracy | ± (3 % rdg + 2 digits) |
| DC Voltage Accuracy | ± (2 % rdg + 2 digits) |
The recommended calibration interval is 12 months.
No, these testers have no field calibration. If they fail performance tests, contact Fluke Service for repair.
Required equipment includes a calibrator with ac/dc voltage range 10 V to 600 V, accuracy ±5%, frequency 50/60 Hz, and resistance range 100 Ω to 9 kΩ.