Fluke 1550C/1555 Insulation Tester Calibration Manual

This document introduces the calibration method of 1550C/1555 insulation tester.


Fluke 1550C/1555 Insulation Tester Calibration Manual - cover page
Brand
Fluke
Category
Other Electronics
Document type
Service Manual
Language
English
Pages
32
File format
PDF
File size
655 KB
Published
15 October, 2023
Updated
15 October, 2023
Model
1550C
MD5 checksum
0BC7EE6E067F60C0B75E9BB033F2A448

About this manual

The document serves as a technical guide for qualified technicians to perform verification tests and calibrate the Fluke 1550C/1555 Insulation Tester. It details required equipment, step-by-step adjustment procedures for high voltage, current, and charge, as well as battery replacement and disassembly instructions.

What's inside

  • Introduction
  • Safety Information
  • Required Equipment
  • Performance Test Procedures
  • Adjustment Procedure
  • Battery Replacement Procedure
  • Disassembly
  • Specifications
  • Principle of Measurement

Specifications

Display Size75 mm x 105 mm
Power Source12 V lead-acid rechargeable battery. 2.6 Ahr
Charger Input85 V to 250 V ac, 50/60 Hz, 20 VA
Operating Temperature-20 °C to 50 °C (-4 °F to 122 °F)
Weight3.6 kg (7.94 lbs.)

Frequently asked questions

What is the warranty period for Fluke products?

The warranty period is three years and begins on the date of shipment.

Can standard calibrators be used for insulation tests?

No, do not attempt to use the 5500A, 5520A, or other standard calibrator for insulation and continuity resistance tests as damage will result.

What is the accuracy of the leakage current measurement?

The range is 1 nA to 2 mA with an accuracy of ±(20 % + 2 nA).