Elektro-Automatik PS 2000 B Laboratory Power Supply User Manual

This manual provides a detailed introduction to the operation of the PS 2000 B Single series laboratory power supply.


Elektro-Automatik PS 2000 B Laboratory Power Supply User Manual - cover page
Brand
ELEKtro-Automatik
Document type
User Manual
Language
German
Pages
20
File format
PDF
File size
750 KB
Published
14 October, 2023
Updated
14 October, 2023
MD5 checksum
1E5E8DF27B15D3178AC0A86923927EE0

About this manual

The Elektro-Automatik PS 2000 B is a compact and robust laboratory power supply designed for use in schools, workshops, and laboratories. It features an isolated design with simple handling and includes built-in protection functions such as overvoltage (OVP) and overcurrent (OCP). The device offers three power classes: 100W, 160W, and 320W.

The manual details the installation process, including connection to a grounded mains supply and fuse replacement. It describes the operation of the front panel controls, such as the Preset button for locking settings and the On/Off key for switching the output. Users can adjust voltage and current limits using rotary knobs, with automatic reduction of one limit if the other is changed to prevent exceeding maximum power. This manual is written in German.

What's inside

  • Einleitung
  • Installation
  • Ansichten und Anschlusselemente
  • Bedienung
  • Anzeige und Fehlermeldungen
  • Tasten
  • Sollwerte einstellen
  • Verhalten des Gerates
  • Weitere Anwendungen
  • Technische Daten

Specifications

Power Classes100W, 160W or 320W
Fuse Type5 x 20mm
InterfaceUSB

Frequently asked questions

What is the maximum power output available for the PS 2000 B series?

The manual states that three power classes are available with outputs of 100W, 160W, or 320W.

How much clearance space is required behind the 320W models for ventilation?

At least 10cm of free space must be left behind the device to ensure sufficient ventilation.

How long does the display remain in fine adjustment mode before switching back to coarse adjustment?

The display automatically switches back to coarse adjustment after 10 seconds if no further setting is made.