Benchmark Scientific Digital Dry Bath II Users Manual

Benchmark Digital Dry Bath II is a highly accurate, microprocessor-controlled dry block heating unit capable of heating one, two or four aluminum blocks or aluminum or glass pans. They provide unsurpassed temperature uniformity and accuracy for heating microtubes, test tubes and other small vessels, including microplates and slides. Each dry bath has a bright green LED display, easy to set temperature and time controls and comes complete with a block lifter.


Benchmark Scientific Digital Dry Bath II Users Manual - cover page
Brand
Benchmark Scientific
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
55 KB
Published
13 October, 2023
Updated
13 October, 2023
MD5 checksum
DB59E3CA961ECCB287F35A6522BB7B5B

About this manual

Benchmark Scientific's Digital Dry Bath II is a microprocessor-controlled heating unit designed to heat one, two, or four aluminum blocks or pans. It features high temperature uniformity and accuracy for applications involving microtubes, test tubes, and other small vessels.

The document provides instructions on setting the temperature using arrow buttons, utilizing the built-in digital timer function which alerts users without stopping heating, and performing user calibration to match certified reference thermometers.

What's inside

  • SET UP and OPERATION
  • Setting the Temperature
  • Using and Setting the Timer
  • User Calibration Function
  • EER Code on the display

Specifications

Temperature range (1 block)Amb.+5C to 150C
Temperature uniformity+/-0.2C
Timer resolution1 minute increments

Frequently asked questions

What does the EER code on the display mean?

The EER code appears if the temperature sensor is shorted or senses an ambient temperature below 0C.

How long must a dry bath equilibrate at a calibration temperature?

The unit must be allowed to equilibrate for 45 minutes at the set calibration temperature.

What happens when the timer function reaches the set time?

The dry bath sounds a beeping alarm and flashes the red Heating lamp, but the heater does not turn off.