This document is the user manual and specifications of NI 5772R, which introduces the features and usage methods of NI 5772R, including signal information, specifications, how to use LabVIEW FPGA Example VI and how to create and run your own LabVIEW project
The document provides a comprehensive guide for the NI 5772R Digitizer Adapter Module, which is designed to work with NI FlexRIO FPGA modules. It covers signal information, component-level intellectual property (CLIP) details, and specifications for analog input channels, digital I/O, and power requirements.
| Analog Input Channels | two analog input (AI) channels |
|---|---|
| Resolution | 12-bit |
| Sample Rate (Dual Channel) | 800 MS/s |
| Sample Rate (Single Channel) | 1.6 GS/s |
| Input Impedance | 50 Ω, single-ended |
| Digital I/O Ports | DIO Port 0 and DIO Port 1 |
| Power Supply (AUX) | +5 V (10 mA maximum) |
The NI 5772R is available in two versions: an AC-coupled version and a DC-coupled version.
To ensure specified EMC performance, the length of all I/O cables must be no longer than 3 m (10 ft).
You must install PXI EMC Filler Panels (National Instruments part number 778700-01) in adjacent chassis slots.