This is a procedural guide for using WalterLorenz® Surgical Assist Arm for robotic-assisted total knee arthroplasty
The document serves as a procedural guide for using the WalterLorenz Surgical Assist Arm, a table-mounted electromechanical device designed for controlled retraction and positioning of instruments during surgery. It details how to attach the arm to standard OR table rails and adjust it for various patient sizes and surgeon preferences.
The guide covers setup procedures, including attaching the arm clamp to the bed rail and orienting the base pole. It also describes common retractors used with the device, such as 90 Degree Hohmann, PCL, and Z retractors, and explains how to assemble universal instrument holders like the Universal Forceps Holder and Flat Universal Instrument Holder.
The device is a table-mounted, bionic electromechanical arm designed for controlled retraction and/or positioning of instruments.
Once unlocked, the device allows it to move with 7 degrees of freedom.
Manual instruments may be connected via one of two methods: a proprietary direct connection or utilizing a Universal Instrument Holder.