This document describes the installation, operation, and maintenance of the T1027 cable resistance temperature sensor. The sensor uses Pt100, Pt1000, and Ni1000 measuring resistors with a measuring range of -40°C to +180°C. The sensor features IP67 protection rating and A, B accuracy class.
This manual provides general instructions, safety warnings, and scope of delivery for the JSP T1027 Cable Resistance Temperature Sensors. It details the application for measuring liquid and gaseous media temperatures using Pt100, Pt1000, or Ni1000 resistors. The document includes dimensional drawings for various versions (S31, S32, S33) and describes the construction with stainless steel or brass stems.
The guide covers installation and commissioning procedures, including fixing the sensor in thermowells and electrical connections for two-wire, three-wire, and four-wire configurations. It also outlines operation and maintenance requirements, recommending periodic calibration checks every 12 months to ensure metrological parameters remain within accuracy classes A or B.
| Measuring resistor | 1x / 2x Pt100, Pt1000 or 1x Ni1000 |
|---|---|
| Measuring range (Pt) | -40 to +180 °C |
| Measuring range (Ni) | -50 to +150 °C |
| Accuracy class | A, B according to EN 60751 |
| Stem material | stainless steel 1.4541, brass |
| Housing protection | IP 67 |
| Recommended calibration period | 12 months |
The Pt sensors measure from -40 to +180 °C, while the Ni sensors measure from -50 to +150 °C.
The manufacturer recommends performing periodic checks of calibration parameters every 12 months.
No, the sensors do not need any maintenance, but it is recommended to check the mounting at preselected intervals.