The DP5G is a member of Amptek's DP5 signal processing family. It is a component in a complete γ-ray spectrometer and is optimized for scintillation spectroscopy.
The Amptek DP5G is a core signal processing component designed for scintillation spectroscopy systems. It functions as a digital pulse processor optimized for detecting ionizing radiation, specifically X-rays and gamma-rays. The device is part of the DP5 family and is intended to be integrated into a complete system that includes a detector module, high voltage supply, and interface circuitry. Amptek offers the DP5G alone or with optional PCG interface modules to create complete systems like the Gamma-Rad5 or TB5.
The manual provides detailed guidance on configuring the DP5G for specific scintillators, such as NaI(Tl), and explains the differences between the 20 MHz and 80 MHz ADC options. It covers the device's physical dimensions, connector types (MMCX for signal input), and power requirements. Additionally, the document includes design details of the analog prefilter and charge amplifier, along with application advice on gain settings and pulse shaping.
| Maximum input count rate (T peak=2.4 μs) | 1.5 x 10^5 cps |
|---|---|
| Throughput | >50% |
| Pulse pair resolving time | 0.25 or 0.5 μs |
| Physical Dimensions | 2.0" x 1.75" |
| Weight | 13 g |
| Nominal Input Power (20 MHz, USB) | +3.3 VDC: 165 mA |
| Input Range | +3 V to +6.4 V |
The DP5G is a component in a complete gamma-ray spectrometer, optimized for scintillation spectroscopy.
No, the DP5G does not contain low voltage power supplies or standard USB and Ethernet interface connectors.
The input is the AC coupled anode output of a PMT, which is a current pulse, not a voltage pulse.
The DP5G is available using either a 20 MHz or 80 MHz ADC.