Rosemount Analytical Model 755EX Explosion-Proof Gaseous Oxygen Analyzer Product Data Sheet

The Model 755EX Explosion-Proof Gaseous Oxygen Analyzer from Rosemount Analytical provides a continuous measurement of gaseous oxygen with the highest accuracy and reliability. The analyzer utilizes a compact paramagnetic sensing unit and offers a wide range of measurement options. It has fast response and linear detector output advantages.


Rosemount Analytical Model 755EX Explosion-Proof Gaseous Oxygen Analyzer Product Data Sheet - cover page
Brand
Rosemount Analytical
Category
Other Electronics
Document type
Product Leaflet
Language
English
Pages
4
File format
PDF
File size
1 MB
Published
29 November, 2022
Updated
28 September, 2023
MD5 checksum
C2CBB853CF7877B244BBBE0EC4288904

About this manual

The Rosemount Analytical Model 755EX is an explosion-proof gaseous oxygen analyzer designed for continuous measurement with high accuracy and reliability in hazardous areas. It utilizes a compact paramagnetic sensing unit mounted on a platinum suspension for shock resistance and offers adjustable response times from 5 to 25 seconds.

What's inside

  • Features
  • Principle of Operation
  • Outline and Mounting Dimensions
  • Physical Specifications
  • Alarm Specifications
  • General Specifications
  • Ordering Information

Specifications

Explosion Proof Case RatingClass 1, Groups B, C, and D, Division 1
Weight118.5 lbs. (53.75 kg)
Response TimeAdjustable 5 to 25 seconds
Reproducibility±0.01% oxygen or ±1% fullscale
Sample Flow Rate50 to 500 cc/min
Ambient Temperature Limits20° to 120°F (-7° to 49°C)
Supply Voltage120 V ±10 V, 50/60 Hz or 240 V ±10 V, 50/60 Hz

Frequently asked questions

What is the measurement principle of the Model 755EX?

The Model 755EX measures the paramagnetic susceptibility of oxygen, which is much higher than other common gases.

Is the Model 755EX approved for hazardous areas?

Yes, it is Factory Mutual Approved for Class 1, Div. 1, Groups B, C, and D.

What are the available range options for the analyzer?

Range options vary from 0 to 1% to 0 to 100% fullscale.