AIM AIM9000E UV Multi-Component Gas Analyzer Product Leaflet

AIM9000E is a UV multi-component gas analyzer, featuring simultaneous measurement of multiple gases, UV emission, continuous spectrum, DOAS technology, simultaneous acquisition of measurement and reference signals, xenon flash lamp, etc. It can measure the concentration of multiple gases and is suitable for monitoring emissions from various boilers, kilns, and incinerators.


AIM AIM9000E UV Multi-Component Gas Analyzer Product Leaflet - cover page
Brand
Aim
Category
Other Electronics
Document type
Product Leaflet
Language
Chinese
Pages
3
File format
PDF
File size
474 KB
Published
16 September, 2022
Updated
28 September, 2023
MD5 checksum
FB483C3F082DBB88EF91D617B5A8D4C6

About this manual

The AIM AIM9000E series UV Multi-Component Gas Analyzer is an online monitoring instrument designed specifically for flue gas and industrial process gases. It utilizes divergent UV continuous spectrum DOAS (Differential Optical Absorption Spectroscopy) technology to simultaneously analyze multiple gases while excluding cross-interference. The analyzer features a touch screen display, a wide measurement range from ppb to percentage levels, and supports both manual and automatic calibration. This manual is written in Chinese.

What's inside

  • Product Features
  • Product Description
  • Industrial Applications
  • Municipal Applications
  • DOAS Measurement Principle
  • Measurement Process
  • Performance Indicators

Specifications

Measurement RangeFrom ppb to percentage %
AccuracyFull scale ±5%
RepeatabilityFull scale ±2%
LinearityFull scale ±2%
Zero DriftLess than full scale ±1% within 30 days
Range DriftLess than full scale ±1% within 30 days
Response TimeLess than 10ms
DisplayTouch screen TFT color LCD, 800x600 resolution
Power Supply220 + 10%V AC, 50/60Hz

Frequently asked questions

What gases can the AIM9000E measure?

The analyzer can measure gases such as O3, NOX, NH3, N2H4, HONO, H2O2, SO2, SO3, H2SO4, H2S, CS2, COS, Cl2, ClO2, HCHO, CH3HCO, C2H5CHO, CH4O, C2H2, BTX, and substituted aromatics.

What is the typical application of the AIM9000E?

It is typically used for emission monitoring in various boilers, kilns, and incinerators.

How is the AIM9000E calibrated?

The instrument has the function of manual or automatic zero and range calibration using cylinder gas.

Download PDF (474 KB)