This document describes the working principle of the 2017 version Delta F series oxygen analyzer, the steps for adding electrolyte, and the operation of power on/off. The oxygen analyzer uses a non-consumable sensor to measure the oxygen concentration in the sample gas.
The Delta F Series Oxygen Analyzer operates on a non-consumable electrochemical principle where oxygen molecules diffuse to a cathode and are reduced, while hydroxide ions are oxidized at the anode to reform oxygen. This process allows for the measurement of oxygen concentration based on the resulting current without consuming the electrode or electrolyte.
The manual details procedures for adding electrolyte, which requires removing the probe and filling the reservoir with E-lectrolyte Blue, followed by a 60-minute rest period before use. It also covers power-on self-diagnosis, calibration using standard nitrogen gas, and storage requirements, including draining the electrolyte for transport and managing the electrolyte level during long-term storage. This manual is written in Chinese.
| Power Supply Voltage | 100-240V AC |
|---|---|
| Electrolyte Volume | 125 ml |
| Rest Time After Electrolyte Addition | 60 minutes |
| Calibration Frequency | Every 12 months |
The analyzer must rest for 60 minutes after adding electrolyte before it can be used.
Delta F recommends calibrating the analyzer every 12 months using known concentration gas.