This guide provides troubleshooting steps, calibration procedures, storage advice, and spare parts information for the Delta F Analyzer.
The document offers detailed instructions for troubleshooting and calibrating the Delta F Analyzer, covering common operational issues, leak detection, and sensor maintenance. It also includes safety information for fuse replacement and shipping procedures.
Key sections include storage conditions, sensor maintenance, a replaceable spare parts list, leak testing methods, basic troubleshooting tables, and specifications for power fuses and voltage requirements.
| Sensor Voltage (S+/-) | 0.7 – 1.3 VDC |
|---|---|
| Heater Voltage (H+/-) | 2.2 – 4.0 VDC |
| Maximum voltage between other wire combinations | < 0.10 VDC |
| AC Power Fuse rating | 2.5 A, 250 VAC, IEC Sheet III, Type T, 5X20 mm |
| DC Power Fuse rating | 1 A, 250 VAC, IEC Sheet III, Type T, 5X20 mm |
| 4‑20 mA Output Fuse | 100 mA fast‑acting, automatically resetting circuit breaker |
The storage temperature should not exceed 70 °C (160 °F).
A 5 × 20 mm, 250 VAC IEC Sheet III Type T fuse rated at 2.5 A is required.
The instrument power must be shut off before removing the fuse.