This manual provides instructions for cleaning and sterilizing the Symphony® and Lactina® breastpump kits, as well as for using and maintaining them.
This manual provides instructions for the Medela Symphony and Lactina breastpump kits. It covers safety precautions, cleaning and sterilization procedures for both hospital and home environments, and detailed assembly steps for the specific parts included in each kit.
The guide explains how to operate the pumps, including the stimulation and expression phases, and provides specific care instructions for the tubing and piston/cylinder assemblies. It also includes information on storing and thawing breastmilk and troubleshooting common issues.
| Autoclave Temperature | 272° F for 3 minutes at 29 PSIG |
|---|---|
| Autoclave Temperature (Alternative) | 250° F for 15-20 minutes at 15 PSIG |
| Boiling Sanitization Time | 10 minutes |
| Battery Charge Time | 12 hours |
| Pumping Time (Charged Battery) | approximately 60 minutes |
All disassembled parts can be autoclaved to a maximum of 272° F for 3 minutes at 29 PSIG or 250° F for 15-20 minutes at 15 PSIG.
Disassemble the parts and sanitize by boiling for 10 minutes.
You should charge the batteries for 12 hours upon receipt of the pump.
The pumping time using a completely charged battery is approximately 60 minutes.