This document is an operating manual for the FLEXIFLO QUANTUM enteral pump. It is intended for enteral use only and should not be used for IV administration. The manual describes the features and characteristics of the product.
The FLEXIFLO QUANTUM is a microprocessor-controlled volumetric pump designed for accurate, controlled enteral feeding. It features automatic priming, an optional hourly flush function, and protection against overdelivery. The device includes various alarm indicators such as OCCLUSION, EMPTY, LOW BATTERY, and DOSE COMPLETE.
| Flow Rate Range | 1 to 300 mL/hr in 1-mL increments |
|---|---|
| Accuracy | ±10% or ±0.5 mL/hr, whichever is greater |
| Mean Occlusion Pressure Limit | 15 psi |
| Power Voltage | 120 VAC, 47-63 Hz, 1 Phase |
| Power Consumption | 4 watts (average) |
| Operating Temperature | 50˚F - 104˚F (10˚C - 40˚C) |
| Battery Type | 2.1 Ah rechargeable sealed lead acid |
No, this device is for enteral use only and not for IV use.
No, the pump features automatic priming so no manual priming is necessary.
The pump triggers a DOSE COMPLETE alarm when the set dose is finished.