This is a user manual for LTV ® Series Ventilators, which introduces how to turn on/off the ventilator, set the ventilation mode, monitor data, etc.
This document serves as a Quick Reference Guide for the CareFusion LTV Series Ventilators, specifically models 900, 950, and 1000. It provides essential instructions for operating the device, including turning the unit on and off, performing checkout tests, and setting up various ventilation modes such as A/C, Assist/Control, SIMV, CPAP, and NPPV.
| Power Source | External AC power adapter or external battery |
|---|---|
| Oxygen Inlet Pressure Minimum | 35 PSIG |
Disconnect the patient from the ventilator, press and hold the On/Standby button for 3 seconds, then press the Silence/Reset button to silence the alarm.
This alarm occurs when the average oxygen inlet pressure is less than the minimum acceptable inlet pressure of 35 PSIG.
Remove the ventilator from service and perform the Calibration procedure.