CURLIN MEDICAL 360-9043-REVA Ambulatory Infusion Systems User Manual

This is an addendum to the PainSmart™ IOD Ambulatory Infusion Systems User Manual, which provides patients, technicians, and caregivers with additional information important to the safe and effective use of the Curlin PainSmart™ IOD ambulatory infusion system.


CURLIN MEDICAL 360-9043-REVA Ambulatory Infusion Systems User Manual - cover page
Brand
Curlin
Category
Other products
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
98 KB
Published
03 August, 2012
Updated
30 September, 2023
MD5 checksum
4CBA689B7648E35880B2FD4049AD71F4

About this manual

This document serves as an addendum to the operator's manual for the Curlin Medical PainSmart™ IOD ambulatory infusion system. It outlines specific modifications intended to improve safety and accuracy, particularly regarding pump handling and testing protocols.

The text elevates two existing cautions to warnings: one advising visual inspection of the pump before use, and another prohibiting dropping or hitting the pump against hard surfaces. If a drop occurs, the pump must undergo a volumetric accuracy test before reuse; failure requires returning the unit to Curlin Medical for evaluation.

What's inside

  • Purpose of this Addendum
  • Caution Bullet Number 2 Warning
  • Caution Bullet Number 13 Warning
  • Procedure for Checking Pumps Between Patients
  • Volumetric Delivery Test

Frequently asked questions

What must be done if the pump is dropped or hit?

The pump must be checked for volumetric accuracy prior to reuse. If it fails the test, it must be returned to Curlin Medical.

When should a new patient check procedure be used?

It is intended for use by technicians in outpatient, Home Care, or Alternate Care settings where anti-siphon valves are not used.