This is an addendum to the user manual of Curlin 2000 series ambulatory infusion system
This document serves as an addendum to the user manual for the Curlin 2000 Plus and 2000 CMS ambulatory infusion systems. It outlines specific modifications to the original manual to address safety concerns regarding pump damage from drops and to improve the accuracy of volumetric delivery testing.
The pump must be checked for volumetric accuracy prior to reuse. If it fails the test, it must be returned to Curlin Medical for evaluation.
Pumps used in outpatient, Home Care, or Alternate Care settings must be checked between patients using the specified procedure.
The passing measurement is between 9.5 g and 10.5 g (equivalent to 9.5 ml to 10.5 ml).