CURLIN MEDICAL 340-9009-REV F User Manual

This is an addendum to the user manual of Curlin 2000 series ambulatory infusion system


CURLIN MEDICAL 340-9009-REV F User Manual - cover page
Brand
Curlin
Category
Other products
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
83 KB
Published
03 August, 2012
Updated
30 September, 2023
MD5 checksum
24490BB63BDF2AFADF35E938C0963592

About this manual

This document serves as an addendum to the user manual for the Curlin 2000 Plus and 2000 CMS ambulatory infusion systems. It outlines specific modifications to the original manual to address safety concerns regarding pump damage from drops and to improve the accuracy of volumetric delivery testing.

What's inside

  • Purpose of this Addendum
  • Caution Bullet Modifications
  • Procedure for Checking Pumps Between Patients
  • Outpatient/Home Care Environment
  • Inpatient/Hospital Environment
  • Volumetric Delivery Test
  • Passing Measurement Criteria

Frequently asked questions

What should be done if the pump is dropped or hit?

The pump must be checked for volumetric accuracy prior to reuse. If it fails the test, it must be returned to Curlin Medical for evaluation.

When must pumps be checked between patients?

Pumps used in outpatient, Home Care, or Alternate Care settings must be checked between patients using the specified procedure.

What is the passing measurement for the Volumetric Delivery Test?

The passing measurement is between 9.5 g and 10.5 g (equivalent to 9.5 ml to 10.5 ml).