Describes routine maintenance of the COMPAT® enteral feeding pump
This document provides routine maintenance instructions for the Novartis Nutrition COMPAT enteral feeding pump. It outlines general cleaning steps to remove spillage from the housing, rollers, control panel, and drop sensor, emphasizing safety precautions such as disconnecting power before cleaning.
The guide also details specific procedures for servicing the pump rotor if it does not rotate freely, including disassembly with an Allen wrench, cleaning, reassembly, and securing the set screw. Additionally, it lists recommended general disinfectants (e.g., Cidex, Isopropyl alcohol) and specific agents for AIDS/hepatitis or tuberculosis disinfection.
No, you should not immerse the pump in cleaning solutions nor autoclave it.
A 3/32" (0.0094") Allen wrench is needed to loosen the set screw and pull the rotor from its shaft.
A 10% concentration of 5.25% sodium hypochlorite (household bleach) is recommended.