This document provides instructions for the installation and program settings of the DualFlo® delivery set. It outlines the installation steps and parameter settings to ensure proper operation and appropriate water delivery. It also provides methods for clearing and resetting parameters.
The document serves as a summary of directions for the Compat DualFlo Enteral Pump (Model #199255). It outlines steps for installing the delivery set, programming pump settings such as flow rate and formula strength, and performing automatic priming. The manual emphasizes verifying tube placement and checking roller clamps before starting feeding.
| Water volume adjustment range | 0-500 mL in 1 mL increments |
|---|---|
| Time interval for water delivery | 2 to 12 hours in 1 hour increments |
| Maximum water delivery rate | 75 mL per hour |
| Programmable flush rate | 375 mL/hr |
Press SELECT until Vol. Delivered is displayed, then press CLEAR.
No, you should not autoclave or immerse the pump in cleaning solutions.