Deltec 5700 CADD-TPN Infusion Pump Quick Reference Card for Clinicians

This document introduces the operation instructions of Stryker Infusor 2000 pump


Deltec 5700 CADD-TPN Infusion Pump Quick Reference Card for Clinicians - cover page
Brand
Deltec
Category
Professional medical equipment
Document type
Quick Start Guide
Language
English
Pages
2
File format
PDF
File size
44 KB
Published
03 August, 2012
Updated
30 September, 2023
MD5 checksum
F9A859623F9BD2F6DCB33F0C41835C70

About this manual

This document serves as a quick reference card for clinicians operating the Deltec Model 5700 CADD-TPN Infusion Pump. It provides step-by-step instructions for stopping and starting the pump, changing batteries, setting reservoir volumes, programming infusion parameters (including taper-up and taper-down periods), and attaching administration sets.

What's inside

  • Pump Operations
  • Stopping the Pump
  • Standard Delivery Procedures
  • Changing the Battery
  • Programming Infusion Parameters
  • Priming the Pump
  • Starting the Pump
  • Alarms and Troubleshooting

Specifications

Power Source Options9-volt battery, Power pack (rechargeable) with a battery adapter, Power pack with AC adapter
KVO Rate5 ml/hr

Frequently asked questions

How do you clear the ML GIVEN value?

Press SELECT MODE until "ML GIVEN" appears on the display, then press SET/CLEAR to clear the value.

What causes a continuous variable-tone alarm with 'LO BAT' remaining on?

The battery is too low to operate the pump; operation stops.