Hanna Instruments HI 83141 Portable pH/mV/ºC Meter Instruction Manual

This document is the instruction manual for the HI 83141 portable pH/mV/ºC meter. The manual provides the necessary information for the correct operation of the meter. If additional technical information is needed, please email [email protected].


Hanna Instruments HI 83141 Portable pH/mV/ºC Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
2
File format
PDF
File size
487 KB
Published
10 July, 2012
Updated
29 September, 2023
MD5 checksum
772CD1326E368BF3B0DE22C41EB115E9

About this manual

This document is the instruction manual for the Hanna Instruments HI 83141 Portable pH/mV/ºC Meter. It provides a functional description of the device's keys and connectors, along with detailed operational guides for measuring pH, mV (ORP), and temperature.

The manual includes specific instructions for calibrating the meter using offset and slope trimmers, replacing the 9V battery, and interpreting low-battery indicators. It also lists technical specifications such as measurement ranges, accuracy, resolution, and environmental limits.

What's inside

  • Warranty
  • Functional Description
  • Specifications
  • Accessories
  • Operational Guide
  • Calibration
  • Battery Replacement

Specifications

pH Range0.00 to 14.00 pH
mV Range±1999 mV
Temperature Range0.0 to 100.0°C
pH Resolution0.01 pH
Temperature Resolution0.1°C
pH Accuracy (@20°C)±0.01 pH
Battery Type1 x 9V alkaline
Battery Life100 hours of continuous use
Operating Environment Temperature0 to 50°C (32 to 122°F)

Frequently asked questions

What is the warranty period for the meter?

All Hanna Instruments meters are warranted for two years against defects in workmanship and materials.

What is the warranty period for the electrodes and probes?

The electrodes and the probes are warranted for a period of six months.

How do I know if the battery is low?

When the battery becomes weak, the meter displays a blinking additional decimal point on the left side of the LCD.

What type of battery does this meter use?

The meter uses one 9V alkaline battery.