Hanna Instruments HI 8043 Portable Dissolved Oxygen Meter Instruction Manual

The instruction manual for the HI 8043 portable dissolved oxygen meter introduces the features and characteristics of the product, including the basic composition, usage methods, maintenance and maintenance.


Hanna Instruments HI 8043 Portable Dissolved Oxygen Meter Instruction Manual - cover page
Brand
Hanna Instruments
Category
Other Electronics
Document type
User Manual
Language
English
Pages
13
File format
PDF
File size
441 KB
Published
10 July, 2012
Updated
30 September, 2023
MD5 checksum
F01D3E5074F96C512452295B743EA2B0

About this manual

The Hanna Instruments HI 8043 is a portable meter designed for measuring dissolved oxygen (DO) in both field applications, such as wastewater treatment and fish farming, and laboratory settings. The instrument features an easy-to-read LCD display and provides readings compensated for temperature effects (ATC). It includes a polarographic probe with a built-in thermistor for simultaneous temperature measurement.

What's inside

  • Preliminary Examination
  • General Description
  • Functional Description
  • Specifications
  • Probe Initial Preparation
  • Dissolved Oxygen Calibration
  • Temperature Calibration
  • Taking Measurement
  • Probe & Membrane Maintenance

Specifications

Battery Type1x9V
Battery Lifeapprox. 100 hours of continuous use

Frequently asked questions

How long should the probe be polarized before taking measurements?

The meter must be switched to the "STB" mode and left for at least 30 minutes to keep the probe continuously polarized.

When should the dissolved oxygen calibration be verified?

Calibration should be checked after approximately 20 hours of use, whenever electrolyte or membrane is changed, or after using aggressive solutions.

What type of battery does the HI 8043 require?

The instrument requires one 9V battery for operation.