Agilent Capillary Isoelectric Focusing Technical Note

Introduces a capillary electrophoresis technology that can be used for protein and peptide separation


Agilent Capillary Isoelectric Focusing Technical Note - cover page
Brand
Capillary
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
188 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
31D368441569A979C61C9F1A239C9767

About this manual

This document provides a technical guide for performing capillary isoelectric focusing (cIEF) using an Agilent capillary electrophoresis system. The method allows for the separation of proteins and peptides based on their isoelectric points (pI), which is useful for identification, purity assessment, and clinical diagnosis.

The protocol details a two-step process: first, focusing proteins into sharp zones within a pH gradient formed by ampholytes, and second, mobilizing the contents past the detector. It includes specific instructions on sample preparation to prevent precipitation, such as adding Triton X-100 and Hydroxyethylcellulose (HEC), along with detailed recipes for preparing reagents like the ampholyte mixture, catholyte, and anolyte.

What's inside

  • Introduction
  • Isoelectric focusing principles
  • pI determination
  • Protein quantitation
  • Sample considerations
  • Method parameters (HPCE mode)
  • cIEF method protocol

Frequently asked questions

What is the recommended protein concentration range for cIEF?

We recommend using a protein concentration range from 0.5 to 2.5 mg/mL.

How should the PVA coated capillary be prepared?

The supplied 64.5 cm x 50 µm i.d. capillary should be cut to a total length of 48.5 cm, providing an effective length of 40 cm.

What are the vial positions for the cIEF method?

Anolyte is position 4, Catholyte is position 5, Acidic wash is position 3, Blank mix is position 2, and Waste (100-µL water) is position 1.