Velocity11 VPrep Liquid Handling Platform Technical Note

This document describes the features and performance of the Velocity11 VPrep liquid handling platform, as well as its application in laboratory automation processes. The VPrep platform offers high precision and accuracy, and can accommodate different types of microplates. It improves sample processing and throughput, while reducing operational errors.


Velocity11 VPrep Liquid Handling Platform Technical Note - cover page
Brand
Velocity11
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
120 KB
Published
05 July, 2012
Updated
29 September, 2023
MD5 checksum
6FEED8FD9DE1675EA7CCECBE3A58D64C

About this manual

The document presents a technical evaluation of the Velocity11 VPrep liquid handling platform, specifically focusing on its performance with a 96 Channel ST Disposable Tip Head. The study measures accuracy and precision across low volume ranges (0.3 to 2.0 µL) using tartrazine/DMSO solutions.

What's inside

  • Introduction
  • VPrep Features
  • Accuracy and Precision Testing
  • Materials
  • Method
  • Results

Frequently asked questions

What is the practical volume range for pipetting with the VPrep and 96 Channel ST Disposable Tip Head?

The system can pipette across a low volume range of 0.3 to 10 uL.

What are the precision and accuracy results for dispensing 2.0 uL?

At 2.0 uL, the precision is 1.4% CV with +1.4% accuracy.

How long does it take to transfer a plate at the tested volumes?

The transfer time is 37 seconds for 0.3 uL and 0.5 uL, or 40 seconds for 2.0 uL.