Agilent Cary 60 UV-Vis Spectrophotometer Baseline Correction Guide

This document introduces the design features of the Agilent Cary 50 and Cary 60 UV-Vis spectrophotometers, including double-beam design and internal reference detector.


Agilent Cary 60 UV-Vis Spectrophotometer Baseline Correction Guide - cover page
Brand
Measuring
Category
Other
Document type
User Manual
Language
English
Pages
4
File format
PDF
File size
158 KB
Published
05 July, 2012
Updated
30 September, 2023
MD5 checksum
30D8A31518925A4F823B3C0B880872B9

About this manual

This document provides a technical overview of the Agilent Cary 60 UV-Vis spectrophotometer, highlighting its dual-beam design with two detectors and internal reference. It explains how the instrument compensates for source intensity fluctuations and solvent/cuvette effects without requiring a reference cuvette.

The guide details the step-by-step procedure for obtaining baseline-corrected spectra using the WinUV software. It includes instructions on performing a zero measurement (blank) and measuring the sample, along with the mathematical equations used by the software to calculate corrected absorbance values.

What's inside

  • Introduction
  • Instrument Design
  • Rear Beam Attenuation
  • Signal Variables
  • Blank Measurement
  • Baseline Measurement
  • Step-by-Step Procedure
  • Mathematical Equations
  • Conclusion

Frequently asked questions

What is the maximum absorbance value for turbid samples that the Cary 60 can measure accurately?

The Cary 60 can measure turbid solutions up to 3.3 Abs accurately because the beam is focused near the sample detector.

Download PDF (158 KB)