Agilent 6820 Gas Chromatograph System Data Sheet, includes product dimensions and weight, ambient operating temperature, ambient operating humidity, storage extremes, operating altitude, safety and regulatory certifications, power requirements etc.
The Agilent 6820 Gas Chromatograph System is designed for indoor use with specific environmental conditions, including an ambient operating temperature range of 5 °C to 45 °C and humidity levels from 5% to 90%. The unit measures 68 cm in width, 50 cm in height, and 52 cm in depth, with an average weight of 60 kg. It is compliant with various international safety and electromagnetic compatibility standards, such as CSA/NRTL 61010A-1 and CISPR 11/EN 55011.
| Average Weight | 60 kg |
|---|---|
| Operating Altitude | 4000m |
| Maximum Power Consumption | 2950 VA |
| Frequency Range | 47.5~63 Hz |
| Column Oven Max Temperature | 425 °C |
| Temperature Setpoint Resolution | 1 °C |
| Maximum Temperature Ramp Rate | 75 °C/min |
| Maximum Run Time | 999.99 min |
| Number of Heated Zones | six |
| Maximum Inlets Installed | two |
| Maximum Detectors Installed | two |
| Data Acquisition Rate (FID) | 200 Hz |
| Operating System | Windows 2000 Professional or Windows XP Professional |
The system supports voltages of 120 V, 200 V, 220 V, 230 V, and 240 V with a frequency range of 47.5~63 Hz.
Available detectors include Flame ionization detector (FID), Thermal conductivity detector (TCD), Electron capture detector (ECD), and Nitrogen phosphorous detector (NPD).
The column oven can operate from 4 °C above ambient up to a maximum of 425 °C.
The system requires Microsoft Windows 2000 Professional (service pack 2) or Windows XP Professional.