Varian Cary 50 UV-Vis Advantage Note 6

This document introduces the features and characteristics of Varian Cary 50 UV-Vis spectrophotometer for measuring microlitre volume samples.


Varian Cary 50 UV-Vis Advantage Note 6 - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
594 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
455C7654E0C68FFD2F6ED71F6EAB84DD

About this manual

This document evaluates the reproducibility and light throughput of the Varian Cary 50 UV-Vis spectrophotometer when fitted with a Hellma Traycell™ ultra-microvolume cuvette. The study aims to demonstrate the system's accuracy in measuring microlitre volumes of chemical and biological samples, such as DNA.

The results indicate that the Cary 50/Traycell system provides significantly higher light throughput compared to competitor systems, resulting in a higher photometric range. Furthermore, the removal and replacement of the Traycell has negligible effects on data, demonstrating excellent reproducibility with a maximum Relative Standard Deviation (RSD) of 1.2% across the 200-1000 nm wavelength range.

What's inside

  • Introduction
  • Summary
  • Materials and Methods
  • Results
  • Discussion and Conclusion
  • Acknowledgements
  • References

Specifications

Wavelength Range200-1000 nm
Maximum RSD1.2%
Sample Volume< 4 pL

Frequently asked questions

What is the maximum reproducibility error for the Cary 50/Traycell system?

The system demonstrates excellent reproducibility with a maximum Relative Standard Deviation (RSD) of 1.2%.

What is the minimum sample volume the Hellma Traycell can measure?

The system is ideal for measuring sample volumes that are less than 4 pL.

What light transmission value was required to align the Traycell?

Adjustments were made to obtain light transmission values greater than 25% at 500 nm.

What is the maximum relative standard deviation observed across the wavelength range?

The results showed a maximum RSD of 1.2% across the wavelength range 200-1000 nm.