Determination of Melamine and Related Compounds Using the Varian 220-MS Ion Trap GC/MS/MS Application Note

This is a document about the method of determining the content of melamine, ammeline, ammelide and cyanuric acid in infant milk-based formula and other food and feed products using the Varian 220-MS Ion Trap GC/MS/MS.


Determination of Melamine and Related Compounds Using the Varian 220-MS Ion Trap GC/MS/MS Application Note - cover page
Brand
Agiletn
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
130 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
CD7BB39C8C14AEC83DF09C50536EC04E

About this manual

The document details an analytical method developed to detect and quantify melamine and related compounds (cyanuric acid, ammeline, and ammelide) in various matrices, including infant milk-based formula, pet food, and other feed products. The method addresses the issue of kidney failure caused by these contaminants.

The procedure utilizes a Varian 220-MS ion trap mass spectrometer coupled with a gas chromatograph (431-GC). It involves sample extraction using diethylamine, water, and acetonitrile, followed by trimethylsilyl (TMS) derivatization. The analysis is performed in MS/MS mode to eliminate matrix interference and achieve quantitative results down to 0.4 µg/g.

What's inside

  • Introduction
  • Instrumentation
  • Reagents and Standards
  • Method
  • Extraction procedure
  • Standard Curve
  • GC Conditions
  • MS Conditions
  • Results and Discussion
  • Conclusion

Specifications

Quantitative range20 to 2000 ppb (0.4 to 40 µg/g)
ColumnVarian FactorFour VF-Xms capillary column, 30 m x 0.25 mm x 0.25 um
Carrier Gas FlowHelium at 1 mL/min
Oven Program75 °C for 1 min to 300 °C at 15 °C/min, and hold 4 min

Frequently asked questions

What is the lowest concentration this method can quantify?

The method can quantitate melamine and related compounds at a concentration as low as 0.4 µg/g.

How does MS/MS mode improve analysis compared to full scan?

In MS/MS mode, interference from the matrix is eliminated, providing extra confidence for identification.

Download PDF (130 KB)