Agilent SEC Fingerprinting of Polyvinyl Alcohols Application Note

This application note describes the use of Agilent PL aquagel-OH columns for the aqueous size exclusion chromatography analysis of polyvinyl alcohols. These columns combine high pore volume and high column efficiency (>35,000 plates/meter) for maximum resolution. The calculated molecular weight averages were compared with manufacturers’ quoted viscosity values. Calibration was done using pullulan polysaccharides.


Agilent SEC Fingerprinting of Polyvinyl Alcohols Application Note - cover page
Brand
Agiletn
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
530 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
CE13D52F2A0EAAE4DAD2E846045B3AA4

About this manual

This document presents an application note by Greg Saunders and Ben MacCreath from Agilent Technologies, Inc. It details the analysis of three polyvinyl alcohol samples using aqueous Size Exclusion Chromatography (SEC). The study utilizes Agilent PL aquagel-OH columns, which combine high pore volume and high column efficiency for maximum resolution. Calibration was performed using pullulan polysaccharides to compare calculated molecular weight averages with manufacturers' quoted viscosity values.

What's inside

  • Introduction
  • Conditions
  • Results and Discussion
  • Conclusion

Specifications

Column Part NumberPL1149-6840
Column Dimensions300 x 7.5 mm
Flow Rate1.0 mL/min
Detection MethodRI
Eluent Composition0.2 M NaNO3 + 0.01 M NaH2PO4

Frequently asked questions

What columns were used for the analysis?

The analysis used 2x PL aquagel-OH 40 8 pm columns.

What was the flow rate used in the experiment?

The flow rate was set to 1.0 mL/min.