Agilent G1978A LC/MS Ion Trap Multimode Source Configuration Guide
This document is the Configuration Guide for the Agilent G1978A Multimode Source designed for Agilent LC/MS Series 6300 ion trap systems. It provides instructions on installing the source, replacing obsolete modules, and converting the system from single-mode sources (ESI, APCI, APPI) to the multimode configuration.
The guide details the installation process, including software setup for Trap Control 6.1 and hardware connections. It also explains how to verify source sensitivity, remove the source, and convert back to single-mode operation. Additionally, it covers creating methods for mixed positive/negative modes and alternating ESI/APCI operations. This manual is written in Italian.
| Product Code | G1978A |
|---|---|
| Compatible System | LC/MS Series 6300 |
| Software Version | Trap Control 6.1 |
| ChemStation Version | B.01.03 or later |
| Required Module | GELV-5 and GEPS-2 |
Obsolete GELV modules must be replaced with GELV-5, and GEPS-1 modules must be replaced with GEPS-2.
The system requires the installation of Trap Control version 6.1.
ChemStation version B.01.03 or later is required for the multimode source.