This document describes the features and characteristics of the INZ-TN513327a-C module, which supports MODBUS communication protocol and is suitable for ZRJ/ZKJ series products.
The Fuji Electric ZRJ/ZKJ gas analyzer features communication capabilities via an RS-232C interface, allowing bidirectional data exchange with a host computer. The system supports a master-slave configuration where one host can connect to one slave directly, or up to 31 slaves when using an RS-232C to RS-485 converter.
This document provides detailed specifications for the MODBUS RTU communication protocol, including wiring diagrams for the CN2 connector, communication parameter settings, and memory mapping for data reading and writing. This manual is written in Chinese.
| Communication Standard | EIA RS-232-C |
|---|---|
| Communication Speed | 9600bps |
| Data Format | 8-bit data, 1 stop bit, no parity |
| Transmission Mode | MODBUS RTU (HEX) |
| Error Check | CRC-16 |
One host can connect to 1 slave directly, or up to 31 slaves when using an RS-485 interface.
A standard RS-232C reverse cable should be used.