The Lumat LB 9507 is a ultra sensitive tube luminometer from BERTHOLD, it provides high sensitivity and versatility, and is ideal for reporter gene assays and other important luminescence-based applications in research and diagnostics.
The Lumat LB 9507 tube luminometer uses an ultra‑fast single photon counting photomultiplier (PMT) with low noise, delivering high sensitivity (1 amol ATP) and a linear dynamic range exceeding six decades. Its measurement chamber holds two tubes in a rotating holder, allowing one sample to be measured while another is loaded, increasing throughput.
The instrument includes variable‑volume injectors (25‑300 µl) based on JET injection technology, built‑in software with multiple measurement protocols (raw data, dual‑label, kinetic, cut‑off, quantitative), and optional WinTerm software for data export via a serial RS‑232 interface. Technical specifications such as power supply (110‑230 V, 50/60 Hz, 60 VA), dimensions (345 × 365 × 180 mm) and operating temperature (15‑35 °C) are provided.
| Spectral range | 380-630 nm |
|---|---|
| Sensitivity | 1 amol ATP |
| Dynamic range | > 6 orders of magnitude |
| Power supply | 110-230 V, 50/60 Hz, 60 VA |
| Dimensions (W×D×H) | 345 x 365 x 180 mm |
| Weight | 14 kg |
| Operating temperature | 15-35 °C |
| Humidity (non‑condensing) | 10-85 % |
The measurement chamber has two tube positions in a rotating holder, allowing one tube to be measured while another is loaded.