This manual provides guidance on using Bioland G-423S blood glucose test strips with compatible meters. It includes storage and handling instructions, testing procedures, and precautions for accurate results.
The test strips are designed for use with the G-423 blood glucose meter only. They must be stored in their original container at 4-40°C (32-104°F), away from light and moisture. Test results may be affected by factors such as improper storage, expired strips, or environmental conditions. This manual is written in Chinese.
| Test Range | 1.1-33.3 mmol/L (20-600 mg/dL) |
|---|---|
| Storage Temperature | 4-40°C (32-104°F) |
| Expiration | 90 days from opening |
The test range is 1.1-33.3 mmol/L (20-600 mg/dL).
Store in the original container at 4-40°C (32-104°F), away from light and moisture.
If the result is too low, the meter displays "L0"; if too high, it displays "HI". Use a new strip to confirm the result.