Fluke 115C and 117C are battery-powered true-rms multimeters with a 6000-count display and an analog pointer display. The manual applies to both models, and all illustrations use the 117C. The meters meet CAT III requirements of IEC 61010-1, second edition.
The manual describes safety precautions, display symbols, input terminals, rotary switch positions, sleep mode, MIN MAX AVG recording, display hold, backlight, manual and auto ranging, and accuracy specifications. The 117C adds Auto Volts, VoltAlert non-contact voltage detection, and a low-impedance voltage measurement mode. This manual is written in Chinese.
| Display | 6000-count display with analog pointer display |
|---|---|
| Safety category | CAT III per IEC 61010-1, second edition |
| Warranty | 3 years |
| Sleep mode timeout | 20 minutes |
| Backlight auto-off | 40 seconds |
| Danger voltage indication | 30 V or greater, or overload |
| Auto Volts low input impedance | Approximately 3 kΩ |
| DC voltage range | 0.001 V to 600 V |
| AC voltage range | 0.06 V to 600 V |
| Resistance range | 0.1 Ω to 40 MΩ |
| Capacitance range | 1 nF to 9999 µF |
| DC current range | 0.001 A to 10 A |
| AC current range | 0.1 A to 10 A |
| Frequency range | 5 Hz to 50 kHz |
| Diode test | Overload above 2.0 V |
| Continuity buzzer | On below 20 Ω, off above 250 Ω |
| Low-impedance capacitance | 1 nF to 500 µF |
It appears briefly when the rotary switch is turned to or away from the A (amps) position, reminding the user to check test lead connections.
It enters sleep mode after 20 minutes without a function, range, or button change.
Hold the g button while turning the meter on.
The backlight turns off automatically after 40 seconds.