Sonbest SLI1 Single Bus Intelligent Temperature Data Logging Device Operating Manual

SLI1 series of single-bus intelligent temperature data logging devices operating manual, the product adopts American DALLAS linear field bus technology, can achieve practical online instrument monitoring scheme in low temperature conditions, can be widely used in various real-time temperature monitoring of data acquisition applications.


Sonbest SLI1 Single Bus Intelligent Temperature Data Logging Device Operating Manual - cover page
Brand
SNB
Category
Industrial Supply
Document type
User Manual
Language
English
Pages
9
File format
PDF
File size
464 KB
Published
16 January, 2012
Updated
30 September, 2023
MD5 checksum
8EF6C5D3FAF69BC56510F49CCE671039

About this manual

The SLI1 series is a single‑bus intelligent temperature data‑logging device that uses Dallas 1‑Wire technology and supports DS18B20 sensors. It is designed for real‑time temperature monitoring in applications such as industrial plants, cold storage, pharmaceutical GMP, and telecommunications rooms. The device can connect up to 24 sensors on a bus up to 200 m long and offers RS232, optional CAN and RS485 communication ports.

The manual describes hardware connections, power requirements, interface pin definitions, and the user interface with four keys for setting passwords, selecting channels, and configuring alarm limits. It also covers wireless/GPRS extensions, sensor test procedures, and a selection guide for different channel configurations.

What's inside

  • Introduction
  • Technical Parameters
  • Typical Program
  • Interface and Pin Definitions
  • Power and Communication Ports
  • Sensor Test Interface
  • Display Panel and Keyboard
  • Alarm and Channel Settings
  • Selection Guide
  • Random List

Specifications

Display temperature range-55°C~+125°C
Display resolution0.1°C
Temperature measurement accuracy±0.5°C
Temperature rate850 ms per point
Number of channels1‑24 points
Baud rate19200 (customizable)
Communication portsRS232 (optional CAN, RS485)
Power supplyDC 5 V 2 A
Power consumption2 W
Weight270 g
Operating temperature range-40°C~+85°C
Storage temperature range-40°C~+85°C
Maximum bus cable length<200 m
Dimensions72 mm × 72 mm × 75 mm
Hole size67 mm × 67 mm

Frequently asked questions

What communication interfaces are supported by the SLI1?

The device includes an RS232 standard serial port and optional CAN or RS485 ports. It also provides a three‑wire bus interface that can reach up to 200 m and support up to 24 sensors.

How many sensors can be connected to the SLI1 bus?

Up to 24 sensors can be connected on the single‑bus interface, with a maximum cable length of 200 m.

Download PDF (464 KB)