Hitachi CR22G/CR21G Error TEMP

High-speed refrigerated centrifuge · taken from the manufacturer’s manual, pages cited

Alarm codeTEMP

TEMP display shows a temperature over 50°C, or a temperature lower or higher than the set temperature more than 20°C. The manual's first action is: “Lower the room temperature if it is over 35°C. Clean the radiator if clogged.”

Model
Hitachi CR22G/CR21G
Display shows
TEMP
Type
Alarm code
Source
Manual p. 55

What TEMP means

TEMP display shows a temperature over 50°C, or a temperature lower or higher than the set temperature more than 20°C. p. 55

Common causes

  • TEMP display shows a temperature over 50°C, or a temperature lower or higher than the set temperature more than 20°C. p. 55

What to do

  1. Lower the room temperature if it is over 35°C. Clean the radiator if clogged. p. 55

Before you service the product, read the safety instructions in the manual p. 3.

If TEMP is still not resolved

  • contact your local dealer or Hitachi Koki service representative p. 3
    “Do not perform any operation not specified in this manual. If any problem is found on your instrument, contact your local dealer or Hitachi Koki service representative.”
  • Contact Hitachi Koki authorized sales/service representative. p. 6
    “Contact Hitachi Koki authorized sales/service representative.”

This page summarises what the manufacturer’s manual says about TEMP; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hitachi. Found a mistake? Report an error.